Parameter Optimization in Voltage Pulse Plethysmography

izvorni znanstveni rad

izvorni znanstveni rad

Parameter Optimization in Voltage Pulse Plethysmography

Vrsta prilog sa skupa (u zborniku)
Tip izvorni znanstveni rad
Godina 2007
Nadređena publikacija IFMBE Proceedings
Status objavljeno

Sažetak

Measurement optimization is being studied when short voltage pulses are stimulating bio-tissue and the transient process is sampled in order to measure the tissue volume changes. The measurement sensitivity depends on the ratio of the sampling instant to the time constant of the transient process, (T/t) and on the ratio of the current sensing resistance to the resistance of the electrode-skin interface (R0/RSX). With variations of R0/RSX and T/t the sensitivity changes from negative to positive values, while it equals zero for certain R0/RSX and T/t ratios. The sensitivity is greater when positive while than negative but it depends on T/t and R0/RSX. For negative sensitivity, T and R0 can be chosen to maximize the sensitivity and minimize its variations.Measurement optimization is being studied when short voltage pulses are stimulating bio-tissue and the transient process is sampled in order to measure the tissue volume changes. The measurement sensitivity depends on the ratio of the sampling instant to the time constant of the transient process, (T/t) and on the ratio of the current sensing resistance to the resistance of the electrode-skin interface (R0/RSX). With variations of R0/RSX and T/t the sensitivity changes from negative to positive values, while it equals zero for certain R0/RSX and T/t ratios. The sensitivity is greater when positive while than negative but it depends on T/t and R0/RSX. For negative sensitivity, T and R0 can be chosen to maximize the sensitivity and minimize its variations.

Ključne riječi

voltage pulse plethysmography; electrode tissue interface; tissue resistance variation